期刊文献+

高效液相色谱-质谱联用测定人血浆中奥美拉唑的浓度 被引量:2

Determination of Omeprazole in Human Plasma by HPLC-MS/MS
下载PDF
导出
摘要 目的建立高效液相色谱-质谱联用法测定人血浆中奥美拉唑的浓度。方法以泮托拉唑为内标,血浆样品经乙腈沉淀、取上清稀释后,进样,经HPLC-MS/MS分析。采用Ultimate XB-C18(2.1 mm×150 mm,3μm),流动相甲醇(0.1%甲酸)∶水(0.1%甲酸)=(70∶30,V/V);流速:0.2m L·min-1,柱温:40℃,电喷雾离子源(ESI),以多离子反应监测方式(MRM)进行正离子监测,奥美拉唑和内标泮托拉唑的定量分析离子对分别为m/z 346.2→m/z 198.0和m/z 384.0→m/z 200.0。结果奥美拉唑的线性范围为5.00-2000.00μg·L-1,定量下限为5.00μg·L-1。方法回收率在96.4%-107.9%,日内和日间精密度(RSD)均小于15%。结论该方法操作简便、灵敏、准确、快速,适用于奥美拉唑的人体药代动力学研究。 OBJECTIVE To establish a high performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS) method for the determination of Omeprazole in human plasma. METHODS Human plasma sam- ples were precipitated with acetonitrile and determined by HPLC-MS/MS using pantoprazole as the internal stand- ard. The separation was carried out on a Ultimate XB-C18 (2. 1 mm x 150mm ,3 μm) with a mobile phase of methanol- water ( including 0. 1% formic acid) (70:30) and the flow rate was 0. 2mL · min -1. ESI was performed in the MRM positive mode with using target ions at m/z 346.2-m/z 198.0 (Omeprazole) and m/z 384. 0--m/z 200. 0 (panto- prazole). RESULTS The calibration curve of Omeprazole was obtained in the range of 5.00 - 2000. 00μg· L-1 with good linearity (r 〉0. 99). The LLOQ of Omeprazole was 5μg · L-1. The method recovery was between 96. 4% and 107.9% ,and the intra and inter day RSD were less than 15%. CONCLUSION This method is convenient, sensitive, accurate and fast, which is suitable for human pharmacokinetic studies of Omeprazole.
出处 《海峡药学》 2016年第1期204-207,共4页 Strait Pharmaceutical Journal
关键词 奥美拉唑 高效液相色谱串联质谱法 血药浓度 Omeprazole Liquid chromatography tandem mass spectrometry Plasma concentration of drug
  • 相关文献

参考文献4

二级参考文献12

共引文献22

同被引文献14

引证文献2

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部