丙泊酚联合瑞芬太尼全身麻醉下颧弓部与额部脑电双频指数的比较 被引量:2

Comparison of Bispectral Index Scores Between the Standard Frontal Sensor Position and an Alternative Zygomatic Position Under General Anesthesia by Propofol Combined with Remifentanil
下载PDF
导出
摘要 目的比较分析神经外科手术中丙泊酚联合瑞芬太尼全身麻醉下颧弓部与额部脑电双频指数(BIS)的相关性。方法选择神经外科手术患者20例,术中采用丙泊酚联合瑞芬太尼靶控输注进行全凭静脉麻醉。每例患者同侧颧弓部和额部分别放置两个BIS电极,采集清醒(t1)、气管插管(t2)、切皮(t3)、手术结束(t4),气管拔管(t5)各时间点BIS值进行分析。结果在t1时颧弓部BIS值为(84.85±9.64),明显低于额部的(88.95±6.42,P〈0.01),而在t2-t5时颧弓部与额部的BIS值比较差异无统计学意义(P〉0.05)。散点图分析显示颧弓部与额部BIS值在麻醉前后具有正相关性(R2=0.892,P=0.000)。结论神经外科手术麻醉期间,当额部电极放置受限时,颧弓部可以作为替代部位放置电极进行监测。 Objective To compare the bispectral index score( BIS) correlationbetween standard frontal sensor position and an alternative zygomatic position under general anesthesia with propofol combined with remifentanil during neurosurgery.Methods Twenty patients undergoing neurosurgery were enrolled. Everyone received total intravenous anesthesia by targetcontrolled infusion of propofol and remifentanil. Two BISTMQuatro sensors mounted on the frontal and zygomatic regions were connected to BIS VistaTMmonitors on each patient during general anesthesia.Data from each position were collected and analyzed at the time of awakening( t1),intubation( t2),incision( t3),the end of surgery( t4),and extubation( t5). Results At t1,zygomatic BIS( 88.95±6.42) was significantly lower than frontal BIS( 84.85± 9.64,P〈 0.01). But during anesthesia,different of BIS value was no statistical significance between the standard frontal position and an alternative zygomatic position( P 〉0. 05).Scatter plot analysis revealed a significant correlation between BIS of frontal positionand that of zygomatic position( R2= 0.892,P =0.000). Conclusion During the anesthesia maintenance period,the zygomatic position can be availably used as an alternative position for monitoring if the operative field renders the standard frontal position unavailable.
出处 《医药导报》 CAS 2016年第4期357-359,共3页 Herald of Medicine
关键词 丙泊酚 瑞芬太尼 脑电双频指数 全身麻醉 神经外科 颧弓 Propofol Remifentanil Bispectral index Anesthesia general Neurosurgery Zygomatic arch
  • 相关文献

参考文献8

  • 1MYLES P S, LESLIE K, MCNEIL J, et al. Bispeetral index monitoring to prevent awareness during anaesthesia: the B- Aware randomised controlled trim [ J ]. Lancet, 2004, 363 (9423) : 1757-1763.
  • 2LIU N, CHAZOT T, HAMADA S, et al.Closed-loop coadmin- istration of propofol and remifentanil guided by bispectral index: a randomized muhicenter study [ J ]. Anesth Analg, 2011,112(3) :546-557.
  • 3DAHABA A A, XUE J X, ZHAO G G, et al. BIS-vista occi- pital montage in patients undergoing neurosurgical procedures during propofol-remifentanil anesthesia [ J ]. Anesthesiology,2010,112(3) :645-651.
  • 4SHIRAISHI T, UCHINO H, SAGARA T, et al.A comparison of frontal and occipital bispectral index values obtained during neurosurgical procedures [ J ]. Anesth Analg, 2004,98 (6) : 1773-1775.
  • 5徐国勋,薛纪秀,刘清海,赵国光,王克杰,魏立民.神经外科手术中枕部与额部脑电双频指数相关性分析的研究[J].临床麻醉学杂志,2009,25(1):5-6. 被引量:5
  • 6LEE S Y, KIM Y S, LIM B G, et al.Comparison of bispectral index scores from the standard frontal sensor position with those from an alternative mandibular position [ J ]. Korean J Anesthesiol,2014,66(4) :267-273.
  • 7NELSON P, NELSON J A, CHEN A J, et al. An alternative position for the BIS-Vista montage in frontal approach neurosurgical cases [ J ]. J Neurosurg Anesthesiol, 2013,25 (2) :135-142.
  • 8KIM Y S, LIMB G, LEE l O. False elevation of bispectral index with a mandibular position in a patient undergoing craniotomy [ J ].J Neurosurg Anesthesiol, 2014,26 (3) : 265- 267.

二级参考文献6

共引文献4

同被引文献20

引证文献2

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部