折反式长波线阵红外传感器装调技术 被引量:2

Alignment Technology of Catadioptric Linear Array LWIR Sensor
下载PDF
导出
摘要 针对折反式长波线阵红外传感器的探测距离远、大口径小焦比、无法凝视成像等结构特点与使用要求,提出了一种高精度、流程化、参量化的装调技术,首先保证主反射镜装调后的面型精度,利用准直光管与三坐标测量机严格量化控制主、次镜之间的光学间隔,然后通过波像差测试手段对主、次镜的相对倾斜与径向偏移进行调校;再利用基准转换的方法,保证主、次镜系统与中继镜组光轴一致;最后装配线阵探测器,利用周扫工装反射镜,进行像面与光轴的调校。文中介绍了装调过程中采用的非球面激光定心、微应力粘接装配、中空主次镜光学间隔调校、中继镜"逐片"定心、线阵探测器周视装调等关键技术,结果表明:运用该技术装调后,探测器MDTD可达到4 K,外场探测距离可达60 km以上,满足设计指标要求。 A high-accuracy, standardize alignment technology is brought up, to adapt to the features and requirements of catadioptric linear array LWIR sensor, which are far-reaching, small F-number, and unable to staring imaged. The technique guarantees the accuracy of surface after alignment in the first place by using collimator and CMM which strictly controls the optical interval between the primary and secondary mirror, and then introduces wave front aberration measurement to adjust the tilt-errors and shift-errors of the primary and secondary mirror. After that, the method of reference conversion is adopted to line up the optical axis of the primary and secondary mirror and relay lens. At last, a linear detector is adopted to make adjustments to image and optical axis by using mirror. The paper introduces the key techniques in the alignment process such as centering by laser, microstress bond, alignment of the optical interval between the primary and secondary mirror, alignment for each lens, linear detector alignment etc. The results indicate that by introducing these technologies, the MDTD of the sensors can reach as high as 4 K, and the outdoor detection distance is more than 60 km. All the parameters meet design requirements.
出处 《光电工程》 CAS CSCD 北大核心 2016年第5期82-87,共6页 Opto-Electronic Engineering
关键词 装调 红外传感器 波像差 光学间隔 中心偏差 alignment infrared sensor wave front aberration optical interval centering errors
  • 相关文献

参考文献7

二级参考文献18

共引文献38

同被引文献10

引证文献2

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部