期刊文献+

利用蒙特卡罗建模及估算儿童X-CT检查受照剂量 被引量:8

Modeling and dose estimation of pediatric X-CT examination using Monte Carlo technique
下载PDF
导出
摘要 在儿科X射线-CT检查迅速增加的当下,尤其对电离辐射敏感的儿童,其所致医疗照射剂量的评估是十分重要和迫切的课题。本文结合已建立的1岁儿童体素体模,利用蒙特卡罗技术建立扫描模型,进行CT检查所致儿童受检者的医疗照射剂量估算。建立了CT单层轴向扫描时对1岁儿童36个器官或组织所产生医疗照射剂量的数据库,利用该数据库,可快速估算1岁儿童受检者接受CT扫描时的医疗照射剂量。并与同样扫描条件下的实验测量值进行了比较,结果显示二者相对偏差在25%以内。 With the rapid development and wide application of X- ray CT examination,estimation of the dose from CT exposures has become a hot research spot,especially for children since they tend to have higher radiation sensitivity than adults. In this work,a CT scanner model was developed using the Monte Carlo technique and was applied to estimate organ doses derived from simulated CT exposures with a 1-year- old voxel phantom developed by our group. As a result,a database including doses to 36 organs and tissues was built for 1- year- old children. This medical exposuredose database provides a method for the estimation of 1- year- old patient doses in a variety of CT examinations. And it has been verified by measurement this method is convenient and reliable.
出处 《辐射防护》 CAS CSCD 北大核心 2016年第3期129-134,共6页 Radiation Protection
基金 国家自然科学基金资助项目(批准号11175122) 上海市放射卫生重点学科建设项目(08 GWZX 0602)
关键词 儿科X-CT检查 CT扫描模型 蒙特卡罗 医疗照射 器官剂量 pediatric CT examination CT scanner model Monte Carlo medical exposure organ dose
  • 相关文献

参考文献13

  • 1ICRP. Radiological Protection in Medicine [ M ]. ICRP Publication 105. Oxford: Pergamon Press, 2008.
  • 2郑钧正,高林峰,卓维海,姚杰,王彬,钱爱君,季贵毅,肖虹.上海市放射诊疗发展趋势与医疗照射防护研究[J].辐射防护,2014,34(5):265-273. 被引量:44
  • 3ICRP. Managing Patient Dose in Computed Tomography[ M]. ICRP Publication 87. Oxford: Pergamon Press ,2001.
  • 4UNSCEAR. Sources and Effects of Ionizing Radiation,Vol. I [ M ]. UNSCEAR 2008 Report. New York : UN ,2010.
  • 5王彬,郑钧正,高林峰,卓维海,姚杰,钱爱君.上海市医用X射线CT的应用频率及其分布研究[J].辐射防护,2013,33(2):65-73. 被引量:30
  • 6郑钧正.医疗照射防护是现代社会必须充分重视与强化的热点课题[J].环境与职业医学,2014,31(10):755-757. 被引量:25
  • 7高林峰,郑钧正,卓维海,王彬,曲良勇,姚杰,钱爱君.医用X射线CT主要检查类型所致成年受检者剂量研究[J].辐射防护,2013,33(2):74-81. 被引量:24
  • 8Neves A,Nunes A, Rufino M, et al. Assessment of pediatric CT exposure in [ J]. Radiation Protection 456 - 462. a portuguese hospital Dosimetry, 2012,151 :.
  • 9Gu J,Bednarz B, Caracappa PF, et al. The development, validation and application of amulti - detector CT(MDCT) scanner model forassessing organ doses to the pregnant patient and thefetus using Monte Carlo simulations [ J ]. Phys Med Biol,2009,54:2 699 -2 717.
  • 10Gu JW, Xu X, Caraeappa pregnant patients from modulation calculated PF,et al. Fetal doses to CT with tube current using Monte Carlo simulations and realistic phantoms [ J ]. Radiation Protection Dosimetry,2013,155:64 - 72.

二级参考文献70

  • 1郑钧正,贺青华,李述唐,岳保荣,王琪,杨霞,张建峰.“九五”期间全国医疗照射水平调查研究[J].中华放射医学与防护杂志,2001,21(S1):8-13. 被引量:13
  • 2郑钧正,贺青华,李述唐,岳保荣,王琪,杨霞,张建峰.我国电离辐射医学应用的基本现状[J].中华放射医学与防护杂志,2001,21(S1):13-20. 被引量:64
  • 3郑钧正.加强医疗照射防护的重点工作[J].中华放射医学与防护杂志,2005,25(2):105-106. 被引量:44
  • 4傅征,梁铭会.数字医学概论[M].北京:人民卫生出版社,2009.
  • 5郑钧正.医学影像数字化所面临新课题应对刍议.世界医疗器械(IMD),2010,.
  • 6郑钧正.放射诊疗基本信息资源的开发与利用.世界医疗器械(IMD),2010,.
  • 7白玫,刘彬,郑钧正.心脏介入患者受照剂量研究[J].介入放射学杂志,2007,16(4):222-225. 被引量:14
  • 8郑钧正 贺青华 李述唐 等.“九五”期间全国医疗照射水平调查研究.中华放射医学与防护杂志,2000,20:2-7.
  • 9郑钧正.x射线诊断和介入放射学的放射防护与安全[M]//郑钧正主编.电离辐射医学应用的防护与安全.北京:原子能出版社,2009.
  • 10Matsumoto K, Jinzaki M, tanami Y, et al. Vitual monochromatic spectral imaging with fast kilovohage switching: improved image quality as compared with that obtained with conventional 120 kVp CT[J]. Radiology, 2011, 259 (1):257 - 262.

共引文献110