摘要
目的探讨急诊室心电监护仪报警的原因并提出护理干预对策。方法选取急诊科2014年1—5月收治的200例使用心电监护仪的患者设为对照组,分析该组患者心电监护仪报警次数及报警原因,并制定护理干预对策,选取急诊科2014年6—10月收治的200例使用心电监护仪的患者设为观察组,比较2组报警的发生情况。结果对照组最常见的报警原因是误报警,其次为导联脱落,然后是导联选择不正确。观察组的误报警率低于对照组,差异有统计学意义(P<0.05)。结论导致急诊室心电监护仪报警的原因较多,制定护理干预对策,可减少不正确报警的发生,提高心电监护仪的规范使用质量。
Abstract:Objective To investigate the cause of ECG monitor alarm in emergency room, and to put forward the countermeasure of nursing intervention. Methods A total of 200 patients using ECG monitor in the emergency department from January to May 2014 were as the control group. ECG monitor alarm number and the reason of alarm were analyzed, the measures of nursing intervention was formulated. Another 200 patients in the emergency department from June to October 2014 were in observation group, and incidence of alarm was compared. Results The most common cause of alarm was false alarm, followed by lead off of electrode, and then incorrect electrode connection. The false alarm rate in the observation group was lower than that in the control group, the difference was statistically significant (P〈0.05). Conclusion Nursing intervention measures can reduce incorrect alarm, and improve quality of ECG monitor.
出处
《中西医结合护理(中英文)》
2016年第6期161-163,共3页
Journal of Clinical Nursing in Practice