摘要 目的对免疫透射比浊法测定血清胱抑素C的方法学性能行初步评价,并与免疫散射比浊法进行比较。方法应用Rohe Modular P全自动生化分析仪与SIEMENS BN ProSpec特定蛋白仪分别对血清胱抑素C进行初步的方法学评价,包括精密度、线性回归试验、回收试验、干扰试验。结果免疫透射比浊法和免疫散射比浊法无论是批内还是批间精密度均符合要求,CV均<4.5%;2种方法从低检测范围到高检测范围,均有着良好的线性回归,斜率均接近于1,截距均接近于0,R2均>0.95;免疫透射比浊法与免疫散射比浊法的回收率均>96%;干扰物试验结果显示血红蛋白、三酰甘油及胆红素均对测定无明显影响。结论免疫透射检验的灵敏度较高,检测范围广,与免疫散射试验的结果基本一致,系统误差小、抗干扰能力强,且更加快速、实用和便利。 Objective To initially evaluate the methodological properties of serum cystatin C by transmitted immunotur- bidimetric assay and compare it with scatter turbidimetrie assay. Methods Methodological evaluation was carried out for the serum cystatin C by Robe Modular P automatic biochemical analyzer and SIEMENS BN ProSpec specific protein instrument, including precision, linear regression test, recovery test and interference test. Results Transmitted immuno- turbidimetric assay and scatter turhidimetric assay both conformed to the requirements of intra-or inter-batch precision, and CV was 〈4.5%; the two methods had good linear regression from the low detection range to the high detection range, the slope was both close to 1, the intercept was both close to 0, and R2 was both〉0.95; the recovery rates of transmitted immunoturbidimetric assay and scatter turbidimetric assay were both〉96%; interferent test results showed that hemoglobin, triglyceride and bilirubin had no significant effects on the determination. Conclusion The sensitivity of immune transmission test is high, the detection range is wide, which is consistent with the result of scatter turbidi- metric assay. The system error is small, the anti-interference ability is strong, and it is more rapid, practical and conve- nient.
出处 《中国现代医生》 2017年第21期96-98,102,共4页 China Modern Doctor
关键词 免疫透射比浊法 免疫散射比浊法 血清胱抑素C 方法学 检验 Transmitted immunoturbidimetric assay Scatter turbidimetric assay Serum cystatin C Methodology Test
  • 相关文献

参考文献8

二级参考文献62

  • 1张汉园.透射免疫比浊法和速率散射免疫比浊法测定C反应蛋白的比较[J].江苏大学学报(医学版),2004,14(4):348-348. 被引量:18
  • 2黄飚,唐炜,金坚.免疫分析中的干扰物及其排除[J].标记免疫分析与临床,2004,11(3):177-179. 被引量:8
  • 3李承彬,张国良.即刻法质控运用错误分析[J].检验医学与临床,2007,4(4):250-251. 被引量:6
  • 4冯仁丰.实有医学检验学,第1版[M].上海:科学技术出版社,1996.483-484.
  • 5Blirup-Jensen S. Protein standardization lit :Method optimization basic principles for quantitative deter- mination of human serum proteins on automated in- struments based on turbidimetry or nephelometry [J]. Clin Chem Lab Med,2001,39(11) : 1098-1109.
  • 6Ledue TB,Collins MF,Ritchie RF. Development of immunoturbidimetric assays for fourteen human serum proteins on the Hitachi 912^TM[J]. Clin Chem Lab Med,2002,40(5) :520-528.
  • 7中国疾病预防控制中心.《全国艾滋病检测技术规范》(修订版)[S].2009.
  • 8Kfieka LJ. Human anti-animal antibody interference in immunological assays[J]. Clin Chem, 1999,45 : 942-- 956.
  • 9Kaplan IV,Levinion SS. When is a heterophile antibody not a hetero- phile antibody? When is it an antibody against a specific immunogen [J]. Clin Chem,1999,45:616--618.
  • 10WS/T403-2012.临床生物化学检验指标分析质量指标[s].中华人民共和国卫生部,2012.