摘要 目的分析三种角膜直径测量方法对角膜水平直径测量的一致性及其在后房型有晶状体眼人工晶状体植入术(IOL)中的应用。设计诊断试验。研究对象行双眼ICL植入的患者23例46眼。方法术前应用量规、IOLMaster及Pentacam眼前节分析系统三种测量方法测量角膜水平直径,比较三种方法测量结果的相关性及一致性。主要指标相关系数、95%的一致性界限(Lo A)。结果量规、Pentacam、IOLMaster测量46眼的角膜水平直径分别为(11.60±0.34)、(11.54±0.32)、(11.94±0.35)mm。量规与Pentacam,量规与IOLMaster,Pentacam与IOLMaster测量结果均高度相关(r=0.895、0.897、0.874,P均=0.000)。量规和Pentacam、量规和IOLMaster、Pentacam与IOLMaster测量结果的95%Lo A分别为(-0.24,0.36)mm,(-0.68,0.01)mm,(-0.70,-0.09)mm。结论量规与Pentacam在有晶状体眼人工晶状体植入术测量角膜水平直径的应用中一致性较好,临床应用中可相互替代。量规与IOLMaster、Pentacam与IOLMaster在临床应用中不可相互替代。 Objective To compare horizontal corneal diameter [white-to-white(WTW) distance] measurements with measuring caliper, IOLMaster, and Pentacam in the patients before ICL implantation. Design Diagnostic test. Participants 23 cases(46 eyes) who underwent ICL implantation. Methods The preoperative horizontal corneal diameter was measured using measuring caliper, IOLMaster,and Pentacam. Analysis of the correlation and agreement were performed. Main Outcome Measures 95% limit of agreement(Lo A).Results The mean WTW distance measurements were(11.60 ±0.34) mm as obtained with measuring caliper and(11.54 ±0.32) mm with Pentacam and(11.94 ± 0.35) mm with IOLMaster. Measuring caliper versus Pentacam, measuring caliper versus IOLMaster, Pentacam versus IOLMaster were highly correlated(r=0.895,0.874, 0.897, respectively. all P=0.000). The 95% Lo A for caliper versus Pentacam, caliper versus IOLMaster, Pentacam versus IOLMaster was(-0.24, 0.36) mm,(-0.68, 0.01) mm,(-0.70,-0.09) mm respectively. Conclusion Measuring caliper and Pentacam can be used interchangeably to measure WTW in clinical practice. Measuring caliper and IOLMaster, Pentacam and IOLMaster cannot be considered interchangeable, when it comes to measure WTW.
机构地区 北京美尔目医院
出处 《眼科》 CSCD 北大核心 2017年第4期252-255,共4页 Ophthalmology in China
关键词 角膜直径 IOLMASTER PENTACAM corneal diameter IOLMaster Pentacam
  • 相关文献

参考文献3

二级参考文献56

  • 1陈琛,刘晓华,林丁,刘志成,张昆亚,宋红芳,王玉慧.虹膜组织力学特性及瞳孔阻滞力定量分析的实验装置设计[J].眼科研究,2006,24(2):206-208. 被引量:4
  • 2Wolffsohn JS, Peterson RC. Anterior ophthalmic imaging [J]. Clin Exp Optom 2006;89(4) :205-214.
  • 3Mukesh BN,Le A,Dimitrov PN,Ahmed S,Taylor HR,McCarty CA. Development of cataract and associated risk factors:the visual impairment project [ J]. Arch Ophthalmol 2006 ;124 ( 1 ) :79-85.
  • 4Stifter E, Sacu S,Benesch T, Weghaupt H. Impairment of visual acuity and reading performance and the relationship with cataract type and density [ J ]. Invest Ophthalmol Vis Sci 2005 ;45(5) :2071-2075.
  • 5Chylack LT Jr,Wolfe JK, Singer DM,Leske MC ,Bullimore MA,Bailey IL, et al. The lens opacities classification system Ⅲ. The Longitudinal Study of Cataract Study Group [ J] .Arch Ophthalmol 1993 ;111(6) :831-836.
  • 6McCarty CA, Mukesh BN, Dimitrov PN, Taylor HR. Incidence and progression of cataract in the Melbourne Visual Impairment Project [J]. Am J Ophthalmol 2003; 136 ( 1 ) : 10-17.
  • 7Jose RM, Bender LE, Boyce JF, Heafley C. Correlation between the measurement of posterior capsule opaciflcation severity and visual fimction testing [ J ] . J Cataract Refract Surg 2005 ;31 (3) :534-542.
  • 8Pavlin CJ, Sherar MD, Foster FS. Subsurface ultrasound microscopic imaging of the intact eye [ J ]. Ophthalmology 1990 ;97 ( 2 ) : 244 -250.
  • 9Stachs O, Schneider H, Stave J, Beck R, Guthoff RF. Three-dimensional ultrasound biomicroscopic examinations for haptic differentiation of potentially accommodative intraocular lenses[ J]. Ophthalmology 2005; 102 ( 3 ): 265 -271.
  • 10Ben-Nun J,Alio JL. Feasibility and development of a highpower real accommodating intraocular lens[ J ]. J Cataract Refract Surg 2005 ; 31 ( 9 ) : 1802-1808.

共引文献25

同被引文献66

引证文献8

二级引证文献27

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部