自动管电流联合迭代算法在颈部CT扫描中的应用研究 被引量:7

Application of automatic tube current modulation combined with iterative reconstruction in CT scanning of neck
原文传递
导出
摘要 目的 探讨自动管电流技术联合迭代算法在颈部CT扫描中的应用价值。方法 连续收集行颈部CT扫描(平扫+增强)的患者80例,按随机数表法分为实验组和对照组,各40例。实验组平扫、增强均采用自动管电流技术+迭代算法重建,对照组平扫、增强均采用固定管电流200 mAs+滤波反投影法重建。对两组平扫、增强图像均进行主观评分、客观评价[甲状腺、斜方肌、胸锁乳突肌、颌下腺、翼外肌、脑组织的噪声值(SD)]。比较两组的有效剂量(E)。结果 实验组平扫、增强的E均低于对照组,差异有统计学意义(t=-2.451、-2.451,P〈0.05)。实验组平扫、增强图像主观评分均高于对照组,差异有统计学意义(Z=-1.969、-2.056,P〈0.05);实验组平扫、增强图像甲状腺、斜方肌、胸锁乳突肌、颌下腺的SD均低于对照组,差异有统计学意义(t=2.400、2.516、2.120、2.411、4.134、4.674、2.711、2.892,P〈0.05)。两组平扫、增强图像翼外肌、脑组织的SD差异均无统计学意义(P〉0.05)。结论 自动管电流技术联合迭代算法在颈部CT扫描中,在提高图像质量的同时,可有效降低辐射剂量。 Objective To explore application of automatic tube current modulation combined with iterative reconstruction in CT scanning (non-enhanced and enhanced) of neck. Methods A total of 80 patients with CT scanning of neck were randomly divided into test group(n=40, automatic tube current modulation,iterative reconstruction) and control group(n=40,200 mAs, filtered-back-projection reconstruction). Subjective scores and objective evaluation of two groups were evaluated. Evaluation included image noise(SD) of thyroid gland, trapezius muscle, sternocleidomastoid muscle, submandibular gland, lateral pterygoid muscle and brain tissue. Effective dose (E) was compared. Results The E of non-enhanced and enhanced scanning in test group was lower than that of control group with the statistically significant difference(t=-2.451, -2.451, P〈0.05). The subjective scores of non-enhanced and enhanced scanning in test group were higher than those of control group with the statistically significant difference (Z=-1.969, -2.056, P〈0.05), while SD of thyroid gland, trapezius muscle, sternocleidomastoid muscle, submandibular gland in test group were lower than that of control group with the statistically significant difference(t=2.400,2.516,2.120,2.411,4.134,4.674,2.711,2.892, P〈0.05). There were no significant differences in SD of lateral pterygoid muscle, brain tissue of non-enhanced and enhanced scanning between two groups (P〉0.05). Conclusions Automatic tube current modulation combined with iterative reconstruction in CT scanning of neck could improve image quality, and reduce effectively radiation dose.
作者 胡仕北 何长久 青浩渺 刘舫 李海烈 夏谦 李德珊 周鹏 Hu Shibei;He Changfiu;Qing Haomiao;Liu Fang;Li Hailie;Xia Qian;Li Deshan;Zhou Peng(Sichuan Cancer Hospital & Institute, Sichuan Cancer Center, School of Medicine, University of Electronic Science and Technology of China, Chengdu 610041, China)
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2017年第11期875-879,共5页 Chinese Journal of Radiological Medicine and Protection
基金 四川省科技计划项目(2017JY0080)
关键词 颈部 辐射剂量 迭代重建 自动管电流技术 Neck Radiation dosage Iterative reconstruction ATCM technique
  • 相关文献

参考文献2

二级参考文献17

  • 1Kerl JM, Schoepf UJ, Zweruer PL, et al. Accuracy of coronary artery stenosis detection with CT versus conventional coronary angiography compared with composite findings from both tests as an enhanced reference standard. Eur Radiol,2011,21 : 1895-1903.
  • 2McCollough CH. Re: Maximizing dose reductions with cardiac CT. Int J Cardiovasc Imaging,2009,25:647.
  • 3Hou Y, Yue Y, Guo W, et al. Prospectively versus retrospectively ECG-gated 256-slice coronary CT angiography: image quality and radiation dose over expanded heart rates. Int J Cardiovasc Imaging, 2012,28 : 153-162.
  • 4Morin RL, Gerber TC, McCollough CH. Radiation dose in computed tomography of the heart. Circulation, 2003, 107 : 917-922.
  • 5Funama Y, Taguchi K, Utsunomiya D, et al. Combination of a low- tube-voltage technique with hybrid iterative reconstruction (idose) algorithm at coronary computed tomographic angiography. J Comput Assist Tomogr,2011,35 :480-485.
  • 6Zhang C, Zhang Z, Yan Z, et al. 320-row CT coronary angiography : effect of 100-kV tube voltages on image quality, contrast volume, and radiation dose. Int J Cardiovasc Imaging, 2011, 27: 1059-1068.
  • 7Leipsie J, Labouuty TM, Heilbron B, et al. Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR Am J Roentgenol,2010, 195:649-654.
  • 8Leipsic J, Labounty TM, Heilbron B, et al. Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study. AJR Am J Roentgeno1,2010,195:655-660.
  • 9Hara AK, Paden RG, Silva AC, et al. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. Am J Roentgenol,2009,193:764-771.
  • 10Hou Y, Liu X, Xv S, et al. Comparisons of image quality and radiation dose between iterative reconstruction and filtered back projection reconstruction algorithms in 256-MDCT coronary angiography. AJR Am J Roentgenol,2012,199:588-594.

共引文献73

同被引文献52

引证文献7

二级引证文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部