微风速标准装置的建立和热线风速仪校准方法的实验研究 被引量:16

Experiment Investigation on Calibration of Hot Wire Anemometer Based on Low Air Speed Reference Facility
下载PDF
导出
摘要 中国计量科学研究院建立了微风速标准装置,基于该标准装置对热线风速仪的校准展开了实验研究。微风速标准装置基于相对法测量原理,实验通过气浮滑车搭载热线风速仪运动,由激光干涉仪测得的距离对时间微分得到标准速度,通过这种方式实现了热线风速仪的校准。滑车的运动速度可以溯源到国家计量院的时间和长度基准;此外,根据热线风速仪传感器的理论模型计算得到了热线风速仪的静态特性方程,比较理论模型与校准结果,发现二者具有良好的一致性。实验结果和不确定度分析表明:微风速标准装置的扩展不确定度U=0.82%(k=2);热线风速仪校准结果的扩展不确定度U=2.42%(k=2)。 The low air speed reference facility is established at NIM of China, based on the device, the experiment on calibration of hot wire anemometer is investigated. The facility is described in detail and the calibration method for hot wire sensor is introduced. Based on the reciprocity principle, the sensor probe is mounted on the carriage that slides with consistent speed and the sliding stroke is measured by laser interferometer accurately during calibration process. The velocity measurement is therefore traceable to the measurement of length and time primary standards, additionally, the parameters of static equation is obtained according to theoretical module of hot wire sensor, the experimental results are consistent with theoretical equation. The experimental results and uncertainty analysis show that the expanded uncertainty of the breeze standard device is U = 0. 82% (k = 2), and the expanded uncertainty of calibration result of hot wire anemometer isU=2.42%(k=2).
作者 崔骊水 李鹏 邱丽荣 李建双 CUI Li-shui1, 2, LI Peng2, QIU Li-rong1 , LI Jian-shuang2(1.Beijing Institute of Technology, Beijing 100081, China; 2.National Institute of Metrology, Beijing 100029, Chin)
出处 《计量学报》 CSCD 北大核心 2018年第3期289-293,共5页 Acta Metrologica Sinica
基金 国家自然科学基金(51575502)
关键词 计量学 热线风速仪 微风速 激光干涉仪 校准 不确定度 metrology hot wire anemometer micro air speed laser interferometer calibration uncertainty
  • 相关文献

参考文献7

二级参考文献75

共引文献47

同被引文献119

引证文献16

二级引证文献47

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部