摘要 目的求解支架扩张过程中的应力、形变分布云图,以及压强-直径变化曲线,获取球囊扩张式支架整体性能数据。方法使用"体积控制"球囊-支架模型,运用有限元分析软件ANYSY进行数值模拟,并与实验数据进行比较。结果支架-球囊模型可以很好地模拟支架扩张全过程;当球囊膨胀至最大直径,支架模型的主体结构与加强筋连接位置会发生塑性形变,而最大应力也分布在连接位置。结论通过有限元分析可以对支架性能进行优化。 Objective To obtain the stress and deformation distribution nephogram,the pressurediameter curve of the stents during stent expansion,and to obtain performance data of balloon-expandable stent.Methods The finite element analysis software ANYSY was used to carry out the numerical simulation and compared with the experimental date by using the volume control model of ballon stent. Results The model of balloon stent could well simulate the expansion process of balloon-expandable stent. When the balloon was inflated to the maximum diameter,main structure and reinforcing connection position of stent model could produce plastic deformation,and the maximum stress was distributed at the connecting position. Conclusions The performance of the scaffold can be optimized by finite element analysis.
作者 石更强 潘肖林 李元 吴山山 SHI Gengqiang;PAN Xiaolin;LI Yuan;WU Shanshan(Shanghai Institute for Minimally Invasive Therapy, School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology, Shanghai 20009)
出处 《北京生物医学工程》 2018年第3期241-245,共5页 Beijing Biomedical Engineering
基金 国家自然科学基金(51475308)资助
关键词 球囊扩张式支架 数值模拟 体积控制 有限元分析 应力分布 balloon expandable stent numerical simulation volume control finite element analysis stress distribution
  • 相关文献

参考文献5

二级参考文献52

  • 1赵玉娟,张继红,蔡伟,吴胜群,姜铁民,李玉明.冠状动脉支架术对血小板激活因子干预作用的观察[J].武警医学,2006,17(5):332-335. 被引量:2
  • 2[1]T.Dnerig,A.Pelton,D.Stockel.An Overview of Nitinol Medical Applications.Materials Science and Engineering A273-275(1999)149-160
  • 3[2]The guidance for Industry and FDA Staff:Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems.2005:9.Fatigue Analysis.10.Accelerated Durability Testing
  • 4[3]EN14299-2004(E):Non active surgical implants-particular requirements for cardiac and vascular implants-specific requirements for arterial stents 7.2.8 Fatigue testing
  • 5[4]ISO/CD 25539(2006):Cardiovascular implants-Endovascular devices-Part 2:Vascular stents 8.6.3.5.2 Stress/strain analyses,8.6.3.5.3 Fatigue safety factor determination,8.6.3.5.4 Fatigue durability testing
  • 6[5]M.Pease and B.Walsh.Nitinol constitutive law comparison for finite-element analysis.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:495-500
  • 7[6]W.J.Harrison and Z.C.Lin.The study of nitinol bending fatigue.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:391-396
  • 8[7]D.Tolomeo,S.Davidson and M.Santinoranont.Cyclic properties of superelastic nitinol:design implications.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:471-476
  • 9[8]C.Kulger,D.Maston and K.Perry.Non-zero mean fatigue test protocol for NiTi.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:409-417
  • 10[9]A.Pelton,X.-Y.Gong,T.W.Duerig "Fatigue Testing of Diamond-Shaped Specimens".Proceedings of SMST2003,Monterey,CA