摘要 使用C8色谱柱,以体积分数为90%的甲醇水溶液为流动相,等度洗脱,多反应离子检测模式,监测离子对为m/z547.7→m/z305.4,建立了一次性使用输液器中偏苯三酸三辛酯(TOTM)溶出量的液相色谱-三重四极杆质谱联用(LC-MS/MS)检测方法.结果表明:TOTM质量浓度为5~150 ng/mL时,线性关系良好(r=0.999 4),检出限为0.5 ng/mL,相对标准偏差均小于2.27%,回收率在98.40%~100.22%之间.该方法操作简便、灵敏、准确、重现性好,适用于一次性使用输液器中TOTM的溶出量分析. The LC-MS/MS method was developed and validated for the determination of the TOTM released from disposable infusion set.A C8column was used with the mobile phase of methanol:water(V/V=90∶10)by isocratic elution.The analyte(m/z547.7→m/z305.4)was monitored under MRM mode.The conclusion indicated that the peak area is linear with its contents in the range of5~150ng/mL(r=0.9994),the detection limit is0.5ng/mL.the RSD are less than2.27%,the recovery ratio is in the range of98.40%~100.22%.This method is convenient,sensitive with good accuracy and precision,so that it can serve well in the quantitative analysis of TOTM released from disposable infusion set.
出处 《河南科技学院学报(自然科学版)》 2018年第6期65-68,共4页 Journal of Henan Institute of Science and Technology(Natural Science Edition)
基金 河南省科技计划(182102210501)
  • 相关文献

参考文献8

二级参考文献71

  • 1奚廷斐,王春仁.邻苯二甲酸二异辛酯增塑聚氯乙烯医疗器械的安全性评价[J].中国医疗器械信息,2005,11(2):27-30. 被引量:28
  • 2宋晓峰,魏光辉.医用塑料增塑剂DEHP安全性研究及对策[J].国外医学(卫生学分册),2005,32(6):358-361. 被引量:37
  • 3徐伟东,聂亮.高效液相色谱法-蒸发光散射检测器测定中/长链脂肪乳注射液中磷脂酰胆碱和磷脂酰乙醇胺含量[J].药物分析杂志,2006,26(5):661-663. 被引量:2
  • 4Kambia K, Dine T, Azar R, et al. Comparative study of the leachability of di (2-ethylhexyl) phthalate and tri(2- ethylhexyl) trimellitate from haemodialysis tubing [J]. Int J Pharm, 2001, 229 (1-2): 139-146.
  • 5Kambia K, Dine T, Gressier B, et al. Evaluation of the direct toxicity of trioctyltrimellitate (TOTM), di (2-ethylhexyl) phthalate (DEHP) and their hydrolysis products on isolated rat hepatocytes [J]. lnt J Artif Organs, 2004, 27 ( 11 ) : 971-978.
  • 6Ito R, Miura N, Iquchi H, et al. Determination of trioctyl trimellitate release from PVC tube by LC-MS/MS [J]. Int J Pharm, 2008, 360 (1-2) : 91-95.
  • 7国家标准化管理委员会.GB/T21928-2008,中国标准书号[S].北京:中国标准出版社,2008.
  • 8DEHP的PVC医疗设备的公告,美国FDA,2002年7月12日.
  • 9Directive 2007/47/EC of the european par!lament and of the council, September 2007.
  • 10Michael A. Babich Ph. D et al. "Review of Exposure and Toxicity data for Phthalate Substitutes", U.S. Consumer Product Safety Commission, January 15, 2010.