期刊文献+

立体定向放射治疗设备安全性检测评价方案及案例分析 被引量:4

Evaluation plan and case analysis on the safety detection of stereotactic radiotherapy equipment
下载PDF
导出
摘要 目的:评价立体定向放射治疗设备的安全性检测,保障立体定向放射治疗设备的安全使用。方法:根据国家相关标准制定立体定向放射治疗设备安全性检测步骤和评价方法,分析评价5种不同品牌的立体定向放射治疗设备的机械精度、输出剂量和射野输出均匀性等指标。结果:经对5种立体定向放射治疗设备检测,其等中心精度、光射野一致性、射野输出均匀性以及机架角度改变情况下的X射线输出剂量存在明显差异;而治疗床、治疗头及机架机械精度、输出剂量以及不同剂量率下X射线输出剂量差异不明显。结论:制定立体定向放射治疗设备安全性检测评价方案,对立体定向放射治疗设备进行安全性检测,可为立体定向放射治疗设备的安全性能提供保障。 Objective: To evaluate the safety detection of stereotactic radiotherapy equipment and ensure the safe application of this kind of equipment. Methods: According to the relevant national standards to formulate the procedure and evaluation method of safety detection of stereotactic radiotherapy equipment, as well as to analyze and evaluate series indexes included mechanical accuracy, output dose and uniformity of field output of stereotactic radiotherapy equipments of 5 kinds of brands. Results: Through detected the five kinds of stereotactic radiotherapy equipments, there were obvious differences in isocentric accuracy, consistency of optical field, uniformity of field output, and the output dose of X-ray when the frame angle was changed. While, there were no significant difference in mechanical accuracy and output dose of treatment bed, treatment head and rack, as well as output dose of X-ray under different dosage rate. Conclusion: The formulation of evaluation scheme of safety detection of stereotactic radiotherapy equipment can implement safety detection for this kind of equipment and can provide guarantee for safe performance of this kind of equipment.
出处 《中国医学装备》 2019年第1期11-15,共5页 China Medical Equipment
基金 国家重点研发计划(2016YFC0106703)"立体定向放射治疗设备的可靠性和安全性评价"
关键词 立体定向放射治疗设备 安全性检测 国产放射治疗设备 进口放射治疗设备 Stereotactic radiotherapy equipment Safety performance testing Domestic radiotherapy equipment Imported radiotherapy equipment
  • 相关文献

参考文献7

二级参考文献24

  • 1周平 俗荣利 等.一种270消色差偏转磁铁的研制.第四届全国医用加速器会议论文集[M].黄山:中国物理学会粒子加速器分会、中国癌症研究基金会,1997.116.
  • 2Lutz W,Winston KR,Maleki N.A system for stereotacticradiotherapy with a linear accelerator. International Journal of Radiation Oncology Biology Physics . 1988
  • 3Hartmann GH,Bauer-Kippes B,Serago CF,et al.Precision anaccuracy of stereotactic convergent beam irradiations from alinear accelerator. International Journal of Radiation Oncology Biology Physics . 1993
  • 4Podgorsak EB,Pike GB,Olivier A,et al.Radiosugery withhigh photon beams: a comparison among techniques. International Journal of Radiation Oncology Biology Physics . 1989
  • 5Serago CF,Lewin AA,Houdek PV,et al.Stereotactic target point verification of an X ray and CT localizer. International Journal of Radiation Oncology Biology Physics . 1991
  • 6Bova FJ,Friedman WA.Stereotactic angiography: an inadequate database for radiosurgery. International Journal of Radiation Applications . 1991
  • 7Yeung D,Palta J,Fontanesi J,et al.Systematic analysis of errors in target localization and treatment delivery instereotactic radiosurgery(SRS). International Journal of Radiation Oncology Biology Physics . 1993
  • 8Engler MJ,cURRAN BH,Tsai JS,et al.Fine tuning of linar accelerator accessories for stereotactic radiotherapy. International Journal of Radiation Oncology Biology Physics . 1994
  • 9Stieler F, Wenz F, Shi M, et al. A novel surface imaging system for patient positioning and surveillance during ra- diotherapy. A phantom study and clinical evaluation[J]. Strahlenther Onkoi,2013,189(11) : 938-944.
  • 10Pallotta S, Russo S, Esposito M, et al. PO-0978: testing the C-RAD catalyst elastic image registration software u- sing a deformable female phantom[J]. Radiother Oncol, 2015,115(Suppll) : S520-521.

共引文献78