摘要
本文主要探讨移液器在检定过程中,水分蒸发现象对移液器检定数据的影响,并在移液器检定要求的实验环境中进行简单的水分蒸发试验,对其变化进行观察,最后浅谈能改善水分蒸发影响的方法。
This paper mainly discusses the effect of water evaporation on the test data of the pipette in the verification process,and carries out a simple water evaporation test in the experimental environment required for the verification of the pipette,observes its changes,and finally discusses the methods to improve the effect of water evaporation.
出处
《计量与测试技术》
2019年第2期41-43,共3页
Metrology & Measurement Technique
关键词
移液器
水分蒸发
影响因素
实测试验
改善方法
pipette
water evaporation
influencing factors
field test