期刊文献+

颅脑外伤早期可植入式无线颅内压监测仪的研制 被引量:2

Development of implantable wireless intracranial pressure monitor for early brain trauma
下载PDF
导出
摘要 目的:为了克服有线颅内压监测仪在使用中由于导线造成的并发症,设计一种可植入式无线颅内压监测仪。方法:该仪器利用电磁感应原理设计,由植入颅脑内的体内传感部分和体外仪表部分组成。体内传感部分通过外科手术的方式以胶囊方式植入颅脑内,主要由电容式传感器和与体外便携式仪表通信的感应耦合电感线圈串联而成。体外仪表部分为手持便携式,可随时读取颅内压力。结果:体外仿真实验证明,该仪器能够满足颅内压力检测的需求。结论:可植入式无线颅内压监测仪可准确测得颅内压力,并减少颅内压监护的并发症,具有广阔的临床应用前景。 Objective To design an implantable wireless intracranial pressure(ICP)monitor to eliminate the complications by the wired monitor.Methods The monitor was developed with the principle of electromagnetic induction,which was composed of two parts of the sensor in vivo and instrument in vitro.The sensor in vivo was implanted surgically with the mode of capsule,which consisted of a capacitive sensor connected with an inductively coupled inducter communicating with the portable instrument in vitro.The portable instrument in vitro could read ICP at any moment.Results The simulation experiment in vitro proved that the monitor met the requirements of ICP detection in vitro.Conclusion The monitor developed can measure ICP accurately with the complications reduced efficiently,and thus is worthy promoting clinically.
作者 赵明 魏秋旭 陈健 陈德勇 王军波 ZHAO Ming;WEI Qiu-xu;CHEN Jian;CHEN De-yong;WANG Jun-bo(Department of Neurosurgery,the Fourth Medical Center,Chinese PLA General Hospital,Beijing 100039,China;State Key Laboratory of Transducers Technology,Institute of Electronics,Chinese Academy of Sciences,Beijing 100190,China;University of Chinese Academy of Sciences,Beijing 100049,China)
出处 《医疗卫生装备》 CAS 2019年第5期21-23,共3页 Chinese Medical Equipment Journal
关键词 颅内压 颅内压监测 无线 颅脑外伤 压力传感器 intracranial pressure intracranial pressure monitoring wireless brain trauma pressure sensor
  • 相关文献

参考文献9

二级参考文献56

  • 1梁玉敏,包映晖,江基尧.颅脑外伤后进展性出血性损伤的研究进展[J].中华创伤杂志,2006,22(2):156-159. 被引量:82
  • 2袁卫,张建奇.基于USB2.0和FPGA的多路数据传输系统设计[J].微计算机信息,2006,22(12Z):144-145. 被引量:2
  • 3高文焕,刘润生.电子线路基础.北京;高等教育出版社,2002.5.
  • 4杨素行等.模拟电子技术基础简明教程.北京:高等教育出版社,2003.2.
  • 5王楚,李椿,周乐柱.电磁学.北京:北京大学出版社,2002.26.
  • 6张福学,1996/1997传感器与执行器大全,1997年
  • 7李 华,MCS-51系列单片机实用接口技术,1993年
  • 8LINX Technologies inc. HP SERIES RECEIVER MODULE DESIGNGUIDE. 1998
  • 9Romner B, Grande PO. Traumatic brain injury: Intracranial pressure monitoring in traumatic brain injury [J]. Nat Rev Neurol, 2013, 9(4): 185-186.
  • 10Brain Trauma Foundation, American Association of Neuro- logical Surgeons, Congress of Neurological Surgeons, et al. Guidelines for the management of severe traumatic brain injury. VI. Indications for intracranial pressure monitoring [J]. J Neurotrauma, 2007, 24(Suppl 1): S37-44.