期刊文献+

成人用力肺活量测定的结束标准研究 被引量:3

A study on the end criteria of forced vital capacity curve in adults
原文传递
导出
摘要 目的探讨成人用力肺活量(FVC)测定的结束标准。方法采用多中心横断面研究,连续纳入2017年1—8月在复旦大学附属中山医院、福建医科大学附属第一医院和内蒙古医科大学第三附属医院用肺量测定法进行肺功能检查,且用力呼气量曲线满足出现呼气平台(呼气流量<0.025 L/s、且持续1 s以上)的成人受试者。按通气功能诊断标准分为肺功能正常组(正常组,610名)、阻塞性通气功能障碍组(阻塞组,536例)和限制性通气功能障碍组(限制组,306例),测定不同组受试者的用力呼气时间(FET);探讨以FET的不同时间段为FVC的结束标准时,阻塞性通气功能障碍诊断的敏感度、特异度和安全性。结果正常组、阻塞组和限制组的FET分别为(4.00±1.07)s、(8.08±1.56)s和(2.97±0.76)s,95%CI分别为3.88~4.12 s、7.02~10.14 s和2.21~3.73 s(F=2263.80,P<0.01)。在限制组,随肺功能损害程度加重,FET缩短;在阻塞组,随肺功能损害程度加重,FET延长(均P<0.05)。以出现呼气平台作为FVC的结束标准时,正常组和限制组不良反应率分别为1.1%和1.3%,均低于阻塞组的17.2%(χ2分别为92.73和48.49,均P<0.05)。在阻塞组,FET达6 s、7 s时不良反应发生率分别为4.1%和3.9%(χ2=0.01,P=0.93);其后随呼气时间的进一步延长,不良反应的发生率明显增加。在阻塞组,与FEV1/FVC相比,用FEV1/FEV7诊断的敏感度为99.25%,高于FEV1/FEV6时97.76%(χ2=4.06,P=0.04);诊断的特异度相同,均为100%。结论FET随肺功能状态改变,以FET≥6 s的固定界值作为成人FVC的结束标准并不合适。对于正常人群和限制性肺疾病患者,应以出现呼气平台作为结束标准;对于阻塞性肺疾病患者,以FET达7 s作为结束标准更为合适。 Objective To explore the end criteria of forced vital capacity(FVC)curve in adults.Methods A multicenter cross-sectional study was performed in Zhongshan Hospital Affiliated to Fudan University,the First Affiliated Hospital of Fujian Medical University,and the Third Affiliated Hospital of Inner Mongolia Medical University from January 2017 to August 2017.A consecutive sample of subjects who completed the spirometry test and FVC curves met end criteria of no volume change(<0.025 L)for≥1 s were qualified in this study.Subjects were divided into a normal group(n=610),an obstructive group(n=536),and a restrictive group(n=306)according to pulmonary function test results.The FET values in different groups were compared.The side effects in the 3 groups and the diagnostic accuracy,specificity and security of different FET in the obstructive group were assessed.Results The FET values of the normal group,the obstructive group,and the restricted group were(4.00±1.07)s,(8.08±1.56)s and(2.97±0.76)s respectively,and the 95%CI of FET in the 3 groups were between 3.88-4.12 s,7.02-10.14 and 2.21-3.73 s(F=2263.80,P<0.01).When the exhalation platform was used as the standard of FVC curve,the adverse reaction rate in the normal group and the restricted group were 1.1%and 1.3%respectively,lower than the rate of 17.2%in the obstructive group(χ2=92.73,χ2=48.49 respectively;all P<0.05).In the obstructive group,7 s as the ending criterion had similar incidence of adverse reactions to 6 s(χ2=0.01,P=0.93).With further extension of expiration time,the incidence of adverse reactions increased significantly.In the obstructive group,the sensitivity of FEV1/FEV7 was 99.25%,higher than that at FEV1/FEV6(χ2=4.06,P=0.04),and the specificity of diagnosis was very similar and 100%.Conclusions FET was variable in subjects with different lung function status.It is not appropriate to use a fixed FET≥6 s as the end criterion of spirometry for adults.For patients with normal lung function or restrictive lung function defect,exhalation platform should be used as the end of exhalation standard.For patients with obstructive lung function defect,an FET of up to 7 s is appropriate.
作者 连宁芳 李丽 杨敬平 杨延杰 胡莉娟 朱蕾 Lian Ningfang;Li Li;Yang Jingping;Yang Yanjie;Hu Lijuan;Zhu Lei(Department of Respiratory and Critical Care Medicine,the First Affiliated Hospital of Fujian Medical University,Fuzhou 350005,China;Department of Respiratory and Critical Care Medicine,Zhongshan Hospital Affiliated to Fudan University,Shanghai 200032,China;Department of Respiratory and Critical Care Medicine,the Third Affiliated Hospital of Inner Mongolia Medical University,Baotou 014010,China)
出处 《中华结核和呼吸杂志》 CAS CSCD 北大核心 2020年第6期520-524,共5页 Chinese Journal of Tuberculosis and Respiratory Diseases
基金 上海市重中之重临床医学中心和重点学科建设(2017ZZ02013) 上海市科委医学引导类科技支撑项目(19411969700)。
关键词 用力呼气量 质量控制 肺量测定法 Forced expiratory volume Quality control Spirometry
  • 相关文献

参考文献4

二级参考文献31

共引文献1644

同被引文献24

引证文献3

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部