期刊文献+

An intraoperative lumbar neurological force monitoring system with high-density flexible pressure sensor array 被引量:3

下载PDF
导出
摘要 In the surgery of lumbar disc herniation(LDH),the nerve root retractor is used to pull the nerve root to prevent damage.The traditional medical nerve root retractor cannot quantify the force on the nerve root.In order to improve the nerve root retractor,this paper proposes an intraoperative lumbar neurological force monitoring system.The core module of this system is the improved nerve root retractor equipped with the high density flexible pressure sensor array.The high density microneedle array and multiple pressure detection units are used in the pressure sensor to realise sensitive pressure monitoring in a narrow surgical operation area.The sensing area is 4 mm×17 mm,including 6 detection units.The sensitivity of sensor is 67.30%/N in the range of 0-5 N.This system is used for in vitro animal experiments,which can continuously detect pressure.
作者 Zhang Qi Zhang Xu Li Caili Liu Jianchao Liu Ming Yuan Fang Chen Hongda 张琪;Zhang Xu;Li Caili;Liu Jianchao;Liu Ming;Yuan Fang;Chen Hongda(State Key Laboratory on Integrated Optoelectronics,Institute of Semiconductors,Chinese Academy of Sciences,Beijing 100083,P.R.China;University of Chinese Academy of Sciences,Beijing 100049,P.R.China;Academy of Advanced Semiconductors and Intelligence Technologies,Haining 314400,P.R.China)
出处 《High Technology Letters》 EI CAS 2020年第4期435-441,共7页 高技术通讯(英文版)
基金 the National Key Technologies R&D Program(No.2016YFC0105604) the National Natural Science Foundation of China(No.61474107).
  • 相关文献

参考文献1

二级参考文献13

  • 1Fritzsche M, Elkmann N, Schulenburg E. Tactile sensing: a key technology for safe physical human robot interaction. In: Proceedings of 6th ACM/IEEE International Conference on Human-Robot Interaction, Lausanne, 2011. 139-140.
  • 2Khasnobish A, Singh G, Jati A, et al. Object-shape recognition and 3D reconstruction from tactile sensor images. Med Biol Eng Comput, 2014, 52: 353-362.
  • 3Haris M, Qu H. A CMOS-MEMS nano-newton force sensor for biomedical applications. In: Proceedings of 5th IEEE International Conference on Nano/Micro Engineered and Molecular Systems, Xiamen, 2010. 177-181.
  • 4Hwang E, Seo J, Kim Y. A polymer-based flexible tactile sensor for normal and shear load detection. In: Proceedings of 19th IEEE International Conference on Micro Electro Mechanical Systems, Istanbul, 2006. 714-717.
  • 5Yang Y J, Cheng M Y, Chang W Y, et al. An integrated flexible temperature and tactile sensing array using PI-copper films. Sensor Actuator A-Phys, 2008, 143: 143-153.
  • 6Cheng M Y, Lin C L, Yang Y J. Tactile and shear stress sensing array using capacitive mechanisms with floating electrodes. In: Proceedings of IEEE 23rd International Conference on Micro Electro Mechanical Systems, Hong Kong, 2010. 228-231.
  • 7Hoshi T, Shinoda H. Robot skin based on touch-area-sensitive tactile element. In: Proceedings of the 2006 IEEE International Conference on Robotics and Automation, Florida, 2006. 3463-3468.
  • 8Xu Z, Ming L, Bo W, et al. A wide measurement range and fast update rate integrated interface for capacitive sensors array. IEEE Trans Circuit Syst I-Regul Papers, 2014, 61: 2-11.
  • 9Wei Y, Torah R, Yang K, et al. Screen printing of a capacitive cantilever-based motion sensor on fabric using a novel sacrificial layer process for smart fabric applications. Meas Sci Technol, 2013, 24: 75104.
  • 10Lei K F, Lee K, Lee M. Development of a flexible PDMS capacitive pressure sensor for plantar pressure measuremen. Microelectron Eng, 2012, 99: 1-5.