摘要
建立测定3-氟-4-溴苯酚毛细管气相色谱法。采用HP-5毛细管色谱柱和氢火焰离子化检测器,对3-氟-4-溴苯酚进行定量分析和分离条件研究。结果表明,此方法在1.0~15.0mg/L范围内,峰面积与进样浓度呈良好的线性关系,线性相关系数为0.9991,加标平均回收率为99.20%,相对标准偏差(RSD)<1.2%,样品中各组分分离效果较好,可用于合成工艺研究及成品质量控制。
To establish a capillary gas chromatography method for the determination of 3-fluorine-4-bromophenol. A HP-5 capillary column and a hydrogen flame ionization detector were used to study the quantitative analysis and separation conditions of 3-fluorine-4-bromophenol. The results showed that the peak area had a good linear relationship with the sample concentration in the range of 1. 0 ~ 15. 0 mg/l,the linear correlation coefficient was 0. 9991,the average recovery rate was99. 20,the relative standard deviation( RSD) < 1. 2,and the separation effect of each component in the sample was good. It can be used in synthesis process research and finished product quality control.
作者
吴芳珍
徐景峰
陈晓燕
Wu Fangzhen;Xu Jingfeng;Chen Xiaoyan(Changzhou Engineering Vocational and Technical College of Inspection and Inspection,Jiangsu Changzhou 213164)
出处
《云南化工》
CAS
2020年第11期73-74,77,共3页
Yunnan Chemical Technology