期刊文献+

电诱发电生理技术评估人工耳蜗术后听阈研究进展 被引量:4

Review of Hearing Threshold Assessment after Cochlear Implant Recipients Using Electrophysiological Methods
下载PDF
导出
摘要 人工耳蜗植入术后听阈的准确评估对于指导调机和观察康复效果有重要意义。主观行为测听是听阈的金标准,但低龄儿童难以配合,常需要电生理技术辅助。耳蜗术后通过耳蜗设备即可实现电诱发电生理测试。寻找一种既便捷又能准确反映听阈的测试是当前研究的重点。本文综述了各类电生理技术用于耳蜗术后听阈评估的研究进展,探讨它们的应用价值与局限性。 Accurate postoperative hearing threshold assessment is important in cochlear implant recipients for mapping implant electrodes and evaluating auditory rehabilitation.Although behavioral measures are regarded as“gold standard”,electrophysiological methods are usually needed for infants and young children who cannot cooperate during audiometric tests.These electrophysiological responses can be evoked electrically through the cochlear device.Nowadays studies are focused on developing an easy and accurate approach for hearing threshold assessment.This paper reviews research on postoperative hearing threshold assessment using electrophysiological methods in cochlear implant recipients,and with discussion on their values and viability.
作者 黄宇童 潘滔 马芙蓉 WONG Yutong;PAN Tao;MA Furong(Department of Otolaryngology,Peking University Third Hospital)
出处 《中华耳科学杂志》 CSCD 北大核心 2020年第6期1096-1101,共6页 Chinese Journal of Otology
关键词 电生理技术 人工耳蜗植入 术后听阈评估 电诱发 Electrophysiological Methods Cochlear Implant Postoperative Hearing Threshold Assessment Electrically Evoke
  • 相关文献

参考文献6

二级参考文献38

  • 1周娜,吴子明,张素珍,郭维维,于黎明,杨伟炎,韩东一.不同刺激模式前庭诱发肌源性电位的反应特性[J].中华耳鼻咽喉科杂志,2004,39(8):483-485. 被引量:41
  • 2李兴启,卢云云.听觉诱发电位(AEP)的神经生物学基础及临床应用(10)[J].听力学及言语疾病杂志,2006,14(3):239-241. 被引量:4
  • 3[1]Lai WK,Dillier N. A simple two-component model of the electrically evoked compound action in the human cochlear.Audiol Neurootol, 2000 ;5: 333~ 345
  • 4[2]Brown CJ, Hughes ML, Luk B, et al. The relationship between EAP and EABR thresholds and levels used to program the nucleus 24 speech processor: data from adults. Ear & Hearing, 2000 ; 21:151 ~ 163
  • 5[3]Shallop JK, Facer GW, Peterson A. Neural response telemetry with the nucleus C124M cochlear implant.Laryngoscope, 1999 ; 109 : 1755 ~ 1759
  • 6[4]Hughes ML, Brown CJ, Abbas PJ, et al. Compariosn of EAP thresholds with MAP levels in the nucleus 24cochlear implant: data from children. Ear & Hearing,2000;21.164~174
  • 7[5]Cullington H. Perliminary neural response telemetry results. British Journal of Audiology, 2000 ; 34 : 131 ~ 140
  • 8Geisler CD,Frishkopf LS,Brown RM.The " early" response to clicks in awake subjects.Q Prog Rep Res Lab Elect,MIT,1957,15:144-148.
  • 9Kraus N,McGee T. The middle latency response generating system.Electroencephalogr Clin Neurophysiol Suppl,1995,44:93-101.
  • 10Ozdamar O, Kraus N. Auditory middle-latency responses in humans.Audiology,1983,22:34-49.

共引文献39

同被引文献22

引证文献4

二级引证文献3