摘要 目的对比分析POCT血糖仪与全自动生化分析仪血糖测定结果。方法选取2018年7月至2019年12月我院内分泌科患者15例,抽取患者清晨空腹静脉血3 mL,用便携式血糖仪(POCT)检测全血样品以测定血糖,用全自动生化分析仪检测分离后的血浆,比较两种仪器血糖测定结果。结果POCT血糖仪检测结果与全自动分析仪检测结果高度相关r=0.992(t=12.56,P>0.05),两种仪器检测结果比较(P<0.05),POCT血糖仪检测结果在规定范围内,无明显偏差。结论POCT血糖仪检测检测结果准确度能够达到临床检测要求,可用于临床检测血糖。 Objective To compare and analyze the blood glucose measurement results of POCT blood glucose analyzer and automatic biochemical analyzer.Methods A total of 15 patients in the department of endocrinology of our hospital from July 2018 to December 2019 were selected.3mL fasting venous blood was extracted in the morning,and whole blood samples were detected by POCT to measure blood glucose.The plasma after separation was detected by automatic biochemical analyzer,and the blood glucose determination results of the two instruments were compared.Results There was a high correlation between the results of POCT blood glucose meter and those of the automatic analyzer(t=12.56,P>0.05).The results of the two instruments were compared(P<0.05).The results of the POCT blood glucose meter were within the specified range without significant deviation.Conclusion The CDDS is a high cost,high automation and the precision of the system operation and maintenance of the system must be carried THE accuracy of POCT blood glucose meter can meet the requirements of clinical detection,and it can be used for clinical detection of blood glucose.
作者 成军家 黎红娟 邵宝琪 黄神贤 CHENG Junjia;LI Hongjuan;SHAO Baoqi;HUANG Shenxian(Department of Laboratory Medicine,People's Hospital of Chancheng District,Foshan Guangdong 528000,China)
出处 《中国医疗设备》 2020年第S02期15-16,19,共3页 China Medical Devices
关键词 POCT血糖仪 全自动生化分析仪 血糖测定 对比分析 POCT blood glucose meter automatic biochemical analyzer blood glucose measurement comparison and analysis
  • 相关文献

参考文献8

二级参考文献62

  • 1胡勤辛,叶雄伟,毕其华.20台POCT血糖仪对照试验及现状调查分析[J].实用糖尿病杂志,2005,13(1):21-23. 被引量:22
  • 2池胜英,袁谦,陈筱菲,陈晓东,胡朋华.医院内快速血糖检测质量管理的实施[J].中华检验医学杂志,2006,29(6):565-566. 被引量:50
  • 3周健,贾伟平.血糖稳定性的意义及临床评估[J].中华医学杂志,2006,86(30):2154-2157. 被引量:74
  • 4中华人民共和国卫生部医政司.全国临床检验操作规程[Z].南京:东南大学出版社,1997.313—327.
  • 5康熙雄.床旁即刻临床检测的规范化趋向[C].中华医学会第七次全国检验医学学术会议,2008:76-78.
  • 6徐建新.POCT技术未来发展预见和市场前景[C].中华医学会第七次全国检验医学学术会议,2008:78-80.
  • 7张同庆.生化分析仪血糖检测结果与POCT血糖结果的比对实验分析[C].中华医学会第七次全国检验医学学术会议,2008:325.
  • 8National Committee for Clinical Laboratory Standards. Ancillary blood glucose testing in acute and chronic care facilities approved guideline C30 - A. Villanov : NCCLS ,2002.
  • 9Barone PW,Strano MS. Single walled carbon nanotubes as report- ers for the optical detection of glucose[J]. J Diabetes Sci Technol, 2009,3(2) : 242-252.
  • 10卫生部办公厅.医疗机构便携式血糖检测仪管理和临床操作规范(试行)[Z],2010-10-30.