期刊文献+

非自动衡器测量不确定度评估及其在符合性判定中的应用 被引量:5

Measurement Uncertainty Evaluation of Non-Automatic Weighing Instrument and its Application in Conformity Assessment
下载PDF
导出
摘要 以OIML R76国际建议中非自动衡器的称量试验为例,列举了2种容易对非自动衡器称量性能做出误判的情形。结合电子天平称量试验的3个案例,分析了修正误差与最大允许误差限较为接近时,直接采用"简单判定原则"作出符合性判定的误判风险,通过对线性载荷误差测量不确定度评估,分别对3个案例给出了95.45%、80.6%和18.1%的符合性概率。实验结果表明,在非自动衡器测量过程中,线性测量载荷误差满足最大允许误差限要求的非自动衡器,仍然可能存在不符合合格判定要求的风险,在法制计量中非自动衡器测量不确定度评估是降低产生误判结果的有效方法。 On determining the weighing performance of a non-automatic weighing instrument( NAWI) from a weighing test result,there are 2 kinds of misjudgment risks if simple acceptance rule is applied when the measurement result is close to the maximum permissible error. Taking the test results of 3 electronic balances of a same type as an example,the measurement uncertainty for the corrected indication error is evaluated,whose influences on the conformity assessment is analyzed. The three kinds of conformance probabilities were 95. 45%,80. 6% and 18. 1%,respectively.The experimental results show that in the measurement process of NAWI,the NAWI whose linear measurement load error meets the maximum permissible error limit may still have the risk of failing to meet the requirements of conformity assessment. In legal metrology,the evaluation of measurement uncertainty of NAWI is an effective method to reduce misjudgment results.
作者 蔡常青 王健 陈杭杭 钟瑞麟 李晓萌 CAI Chang-qing;WANG Jian;CHEN Hang-hang;ZHONG Rui-lin;LI Xiao-meng(National Institute of Metrology,Beijing 100029,China)
出处 《计量学报》 CSCD 北大核心 2021年第1期59-65,共7页 Acta Metrologica Sinica
基金 国家重点研发计划(2016YFF0200103)。
关键词 计量学 质量计量 非自动衡器 不确定度 符合性判定 metrology mass measurement non-automatic weighing instruments uncertainty conformity assessment
  • 相关文献

参考文献5

二级参考文献33

  • 1陈继锋,胡清平.误判概率及其在可靠性分析中的应用[J].计量学报,2009,30(4):381-384. 被引量:2
  • 2许少华,刘扬,何新贵.基于过程神经网络的水淹层自动识别系统[J].石油学报,2004,25(4):54-57. 被引量:25
  • 3邓韧,李著信,樊友洪.基于多种正交基函数的模块化过程神经元网络[J].计算机工程与应用,2005,41(30):69-71. 被引量:2
  • 4钟诗胜,朴树学,丁刚.改进BP算法在过程神经网络中的应用[J].哈尔滨工业大学学报,2006,38(6):840-842. 被引量:18
  • 5Marx K A.Quartz Crystal Microbalance:A useful tool for studying thin polymer films and complex biomolecular systems at the solution-surface interface[J].Biomacromolecules,2003,4 (5):1099-1120.
  • 6Johannsen K,Page D,Roy S.A systematic investigation of current efficiency during brass deposition from a pyrophosphate electrolyte using RDE,RCE,and QCM[J].Electrochimica Acta,2000,45:3691-3702.
  • 7Chen Y,Huang X H,Shi H S,et al.A novel and cost-effective method for early lung cancer detection in immunized serum[J].Asian pacific journal of cancer prevention,2011,12(11):3009-3012.
  • 8Chen Y,Huang X H,Shi H S,et al.Rapid detection of ovarian cancer from immunized serum using a quartz crystal microbalance immunosensor[J].Asian pacific journal of cancer prevention,2012,13(7):3423-3426.
  • 9Du B,Johannsmann D.Operation of the quartz crystal microbalance in liquids:derivation of the elastic compliance of a film from the ratio of bandwidth shift and frequency shift[J].Langmuir,2004,20(7):2809-2812.
  • 10Sauerbrey G.Messung von plattenschwingungen kleiner amplitude durch lichtstrommodulation[J].Zeitschrift for Physik A Hadrons and Nuclei,1964,178:291-295.