聚乙二醇凝聚重量法测定硅酸盐中二氧化硅 被引量:1

Determination of Silica in Silicate by the Method of the Polyethylene Glycol Coagulation Weight
下载PDF
导出
摘要 用动物胶凝聚重量法测定硅酸盐岩石中的二氧化硅含量,但是此方法分析流程长,凝聚条件苛刻且难以控制,凝聚程度不够完全,并且滤液中常含有少量杂质使分析结果偏低。通过氢氧化钠熔样,采用聚乙二醇凝聚硅酸盐中SiO2,使分析流程进一步简化,分析条件容易控制,结果准确可靠,满足一般硅酸盐、硅质岩中SiO2分析要求。相对标准偏差<1%。经国家一级标准物质验证,相对误差<0.20%,精密度允许限Yb符合规范要求。 Animal gel agglomeration method as the national standard method is commonly used to determine silica(SiO2)in silicate,but its analysis process is long,the agglomeration conditions were harsh and difficult to control,the degree of agglomeration is not complete,and the filtrate often contains a small amount of SiO2 to make the analysis result is low.In research,polyethylene glycol is used to condense SiO2 in silicate through sodium hydroxide fusion sample.This analysis process is further simplified,the analysis conditions are easy to control,and the results are accurate and reliable to meets the requirements of SiO2 analysis in general silicate and siliceous rock.The relative standard deviation(RSD)is less than 1%.It has been verified by national first-level standard substance that the relative error(RE)is less than 0.20%and the allowable precision limit Yb meets the requirements.
作者 赵前礼 任风利 霍东婵 易忠 熊金平 马雯姗 Zhao Qianli;Ren Fengli;Huo Dongchan;Yi Zhong;Xiong Jinping;Ma Wenshan(Laboratory of Fourth Geological Brigade of Hubei Geological Bureau,Xianning,Hubei 437100)
出处 《资源环境与工程》 2020年第S02期174-176,共3页 Resources Environment & Engineering
关键词 聚乙二醇 硅酸盐 二氧化硅 精密度允许限Yb polyethylene glycol silicate silica the allowable precision limit Yb