US-CT影像融合容积导航技术可重复性的实验研究 被引量:4

Experimental study on repeatability of volume navigation technology based on US-CT image fusion
下载PDF
导出
摘要 目的探讨US-CT影像融合容积导航技术在肝脏体模应用中的价值。方法采用包含3条肝静脉及6枚囊肿模型的体外肝脏体模,应用融合导航定位系统将其CT薄层扫描图像导入超声检查仪虚拟导航系统进行影像融合,在囊肿模型的最大截面层,应用matlab软件计算影像重合率。实验共进行2次:1)2名不同资质的操作者应用影像融合容积导航技术,分别在1天内各自独立完成3次对6枚囊肿模型的融合对位;2)1周后重复上述步骤。结果高年资操作者和低年资操作者都能获得较高的影像重合率和较低的变异系数。2次实验中,2名操作者组间影像重合率明显相关;不同操作者,前后两次组内实验影像重合率明显相关。结论US-CT影像融合容积导航技术可以达到较高的影像重合率,在组间及组内都具有较高的可重复性。并且,不同操作经验的医师只要在标准操作的前提下,不会显著影响影像重合率的可重复性。 Objective Methods Three hepatic veins and several cyst models were included in the liver phantom.The three hepatic veins and cyst models were clearly visible in CT images and conventional ultrasound scanning.Two different experienced operators performed three times of the image fusion process on the phantom at the same day in one week and repeated the process one more time on the next week,and six cysts were selected to record the fusion counterpoint rate for each time.Results By comparing the data between groups and within groups,we found no significant difference in the results.Conclusion US-CT image fusion volume navigation technology is feasible,repeatable,and highly accurate,which can be used as an effective guidance method in the process of ultrasound interventional diagnosis and treatment.
作者 张黎明 宗维 荆雪红 王刚 丛林 赵斌 ZHANG Liming;ZONG Wei;JING Xuehong;WANG Gang;CONG Lin;ZHAO Bin(Department of Thoracic Surgery,The Second Hospital of Shandong University,Jinan 250033,P.R.China;Shandong Institute of Medical Imaging,Affiliated to Shandong University,Jinan 250021,P.R.China)
出处 《医学影像学杂志》 2021年第3期513-516,共4页 Journal of Medical Imaging
关键词 影像融合容积导航技术 影像重合率 可重复性 Multimode image fusion volume navigation technology Overlap rate of images Repeatability
  • 相关文献

参考文献4

二级参考文献48

  • 1许文哲,李福元,刘强,李成利.不同影像技术导引纵膈病变穿刺活检术评价[J].医学影像学杂志,2012,22(2):265-267. 被引量:4
  • 2Molloy JA,Oldham SA.Benchmarking a novel ultrasound-CT fusion system for respiratory motion management in radiotherapy:assessment of spatio-temporal characteristics and comparison to 4DCT.Med Phys,2008,35:291-300.
  • 3Ewertsen C,Grossjohann HS,Nielsen KR,et al.Biopsy guided by real-time sonography fused with MRI:a phantom study.Am J Roentgenol,2008,190:1671-1674.
  • 4Kliewer MA,Sheafor DH,Paulson EK,et al.Percutaneous liver biopsy:a cost-benefit analysis comparing sonographic and CT guidance.Am J Roentgenol,1999,173:1199-1202.
  • 5Shin HJ,Amaral JG,Armstrong D,et al. Image-guided percutaneous biopsy of musculoskeletal lesions in children.Pediatr Radiol,2007,37:362-369.
  • 6Kang M,Kalra N,Gulati M,et al.Image guided percutaneous splenic interventions.Eur J Radiol,2007,64:140-146.
  • 7Ewertsen C.Image fusion between ultrasonography and CT,MRI or PET/CT for image guidance and intervention-a theoretical and clinical study.Dan Med Bull,2010,57:B4172.
  • 8Livraghi T, Solbiati L, Meloni MF, et al. Treatment of focal liver tumors with percutaneous radio-frequency ablation: complications encountered in a multicenter study [J]. Ra- diology, 2003,226(2) : 441-451.
  • 9Yau T,Tang VY,Yao TJ,et al. Development of Hong Kong Liver Cancer staging system with treatment stratifi- cation for patients with hepatocellular carcinoma [J]. Gas- troenterology, 2014,146(7) : 1691-1700.
  • 10Korean Liver Cancer Study Group (KLCSG),National Cancer Center,Korean (NCC). 2014 KLCSG-NCC Korean Practice Guideline for the management of hepatocellular carcinoma [J]. Gut Liver, 2015,9(3) :267-317.

共引文献16

同被引文献37

引证文献4

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部