摘要 介绍了在组织某次比对活动中,水中^(239)Pu样品的制备、液闪测量方法建立、均匀性检验和结果不确定度评估。液闪测量^(239)Pu的最优PSA(脉冲形状分析,Pulse Shape Analysis)值为62,探测效率约为99.6%。比对样品的均匀性检验采用液闪测量方法,方差分析数据显示液闪测量方法重复性良好,样品通过均匀性检验。不确定度评估结果显示,样品活度浓度的不确定度主要来源于样品均匀性,其不确定度分量占比约96%。 In a comparison activity,the preparation of^(239)Pu aqueous samples,establishment of liquid scintillation counting measurement methods,homogeneity test and uncertainty evaluation are introduced in the paper.The optimal PSA value for LSC is 62,and the detection efficiency is about 99.6%.The homogeneity test uses LSC measurement method.ANOVA data shows that the repeatability of LSC method is good and the samples are uniform.The result of uncertainty evaluation shows that the uncertainty of sample activity concentration is mainly from the homogeneity of the samples.The uncertainty component of the homogeneity is about 96%.
作者 李晓芸 娄海林 李爱云 潘竞舜 卢瑛 文富平 LI Xiao-yun;LOU Hai-lin;LI Ai-yun;PAN Jing-shun;LU Ying;WEN Fu-ping(China Institute of Atomic Energy,Beijing 102413,China)
出处 《核电子学与探测技术》 CAS 北大核心 2020年第6期907-911,共5页 Nuclear Electronics & Detection Technology
基金 食品中生物性及放射性危害物高效识别与确证关键技术及产品研发(2018YFC1602500)。
关键词 ^(239)Pu 液闪测量 均匀性检验 不确定度评估 比对 ^(239)Pu Liquid Scintillation Counting Measurement Homogeneity Test Uncertainty E-valuation Interlaboratory Comparison
  • 相关文献

参考文献3

二级参考文献16

  • 1曹宏燕,钱家松,王成章,蔡填云.标准物质均匀性检验和统计方法研究[J].分析科学学报,1995,11(3):51-54. 被引量:12
  • 2艾承锦,彭明婷,杨雁华.标准物质的均匀性评价[J].中国医药导刊,2007,9(2):155-157. 被引量:18
  • 3Varlam C, Stefanescu I, Duliu O G, et ah Applying direct liquid scintillation counting to low level tritium measure- ment [ J ]. Applied Radiation and Isotopes,2009,67 (5) : 812-816.
  • 4Nikolov J, Todorovic N, Jankovic M, et al. Different meth- ods for tritium determination in surface water by LSC [ J ]. Applied Radiation and Isotopes ,2013,71 ( 1 ) :51-56.
  • 5Reddy P J, Bhade S P D, Narayan K K, et al. Compara- tive study of different methods for the activity quantifica- tion3H and14C radionuclides in dual labeled sample su- ing liquid scintillation analyzer [ J ]. Applied Radiation and Isotopes,2009,67(10) :1945-1951.
  • 6Pates J M ,Mullinger N J. Determination of 222Rn in fresh water:Development of a robust method of analysis by α/ β separation liquid scintillation spectrometry [ J ]. Ap- plied Radiation and Isotopes, 2007,65 ( 1 ) : 92 - 103.
  • 7Bianchi D, Cortese P, Dellacasa G, et al. Determination of the isotopic ratio 234 U/238U in liquid Scintillation and en- vironmental applications [ J]. Applied Radiation and Iso- topes ,2005,553 ( 3 ) :543-549.
  • 8Nebelung C, Baraniak L. Simultaneous determination of 226 Ra,233U and 237 Np by liquid scintillation spectrometry [ J ]. Applied Radiation and Isotopes, 2007,65 ( 2 ) : 209-217.
  • 9Ross H, Noakes J E, Spaulding J D. Iiquid scintillation counting and organic scintillators [ M ]. Michigan, USA : Lewis Publishers, 1989.
  • 10Michael EL’Annunziata.放射性分析手册[M].2版.《放射性分析手册》(第2版)翻译组,译.北京:原子能出版社,2006.