呼气末CO_(2)测量结果偏低的影响因素分析

Analysis of Influencing Factors for Low ETCO_(2) Measurement Results
下载PDF
导出
摘要 该研究介绍了呼气末CO_(2)(ETCO_(2))测量技术的重要性及其在临床使用中的指导意义,重点阐述了深圳迈瑞生物医疗电子股份有限公司(以下简称迈瑞公司)旁流式ETCO_(2)测试模块测量的技术原理,并针对手术室医护人员反馈的ETCO_(2)测量结果偏低的问题,结合产品的设计原理及实际临床场景进行了原因分析,并提出了相关纠正措施,且经过为期3个月的现场调研及随访,临床端反馈纠正措施有效,为使用迈瑞公司ETCO_(2)测量技术的手术室医护人员排查ETCO_(2)测量结果偏低相关问题提供了重要参考,对患者进行有效的ETCO_(2)参数监测及医师参考该结果进行相关病症的准确诊断具有重要的意义。 This study introduced the importance of end-tidal carbon dioxide(ETCO_(2))measurement technology and its directive significance in clinical use,mainly illustrated the technical principles of side-steam ETCO_(2) measurement technology of Shenzhen Mindray Bio-Medical Electronics Co.,Ltd(hereafter refined to as Mindray Corporation).In terms of the low ETCO_(2) measurement feedback from medical staff,depending on the design principles of the product and actually clinical scenarios,the causes were analyzed and the relative corrective measures were put forward.After on-site investigation and follow-up for three months,the corrective measures were turned out to be effective.The study provided an important reference for troubleshooting related problems with Mindray side-steam module low ETCO_(2) measurement results for medical staff and it was of great significance for effective ETCO_(2) monitoring on patients and accurate diagnosis referring to the ETCO_(2) results for doctors.
作者 刘立汉 蔡林意 刘昂驹 刘振兴 Liu Lihan;Cai Linyi;Liu Angju;Liu Zhenxing(Customer Service Department for PMLS Product,Shenzhen Mindray Bio-Medical Electronics Co.,Ltd.,Shenzhen Guangdong 518057,China;Medical Equipment Department,The First People's Hospital of Huaihua City,Huaihua Hunan 418000,China)
出处 《医疗装备》 2021年第15期32-34,共3页 Medical Equipment
关键词 呼气末CO_(2) 旁流式ETCO_(2)测量技术 ETCO_(2)测量结果偏低 原因分析 纠正措施 End-tidal carbon dioxide Side-steam ETCO_(2) measurement technology Low ETCO_(2)measurement results Cause analysis Corrective measures
  • 相关文献

参考文献4

二级参考文献30

  • 1黄马玉.呼气末二氧化碳监测在基层急诊ICU中应用80例体会[J].基层医学论坛,2006,10(4):332-333. 被引量:4
  • 2陈卉.Bland-Altman分析在临床测量方法一致性评价中的应用[J].中国卫生统计,2007,24(3):308-309. 被引量:217
  • 3招伟贤.P(a-et)CO2的变异及其临床意义(综述).2006中华医学会全国麻醉学术年会论文汇编.广州中医药大学第二附属医院.
  • 4Bhende MS. End-tidal carbon dioxide monitoring in pediatrics: concepts and technology. J Postgrad Med 2001;47:153-6.
  • 5Eskaros SM,Papadakos P,Lachmann B.Chapter 44,Respiratory monitoring[M] //Miller RD,Eriksson LI,Fleisher LA,et al.Miller's Anesthesia.7th Edition.Elsevier,2009.
  • 6Raheem MSA,Wahba OM.A nasal catheter for the measurement of end-tidal carbon dioxide in spontaneously breathing patients:a preliminary evaluation[J].Anesth Analg,2010,110(4):1039-1042.
  • 7Soto RG,Fu ES,Vila H Jr,et al.Capnography accurately detects apnea during monitored anesthesia care[J].Anesth Analg,2004,99(2):379-382.
  • 8Deitch K,Miner J,Chudnofsky CR,et al.Does end tidal CO《,2》 monitoring during emergency department procedural sedation and analgesia with propofol decrease the incidence of hypoxic events? A randomized,controlled trial[J].Ann Emerg Med,2010,550):258-264.
  • 9Sandlin D.Capnography for nonintubated patients:the wave of the future for routine monitoring of procedural sedation patients[J].J Perianesth Nuts,2002,17(4):277-281.
  • 10Nishiyama T,Hanaoka K.Usefulness of a new nasal cannula with an end-tidal carbon dioxide measuring port[J].Masui,2003,2(4):424-426.

共引文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部