期刊文献+

电感耦合等离子体发射光谱法测定锌铝镁合金中铝、镁含量 被引量:1

Determination of Al,Mg Contents in Zinc Aluminum Magnesium Alloy by Inductively Coupled Plasma Optical Emission Spectrometry
下载PDF
导出
摘要 采用基体匹配法绘制校准曲线和镧元素作内标校正来消除基体效应和仪器漂移的影响,建立了电感耦合等离子体发射光谱仪测定锌铝镁合金中铝、镁元素的分析方法.将方法应用于锌铝镁合金试样的测定,结果的相对标准偏差(RSD)在0.86%~2.13%之间,加标回收率为96.4%~99.6%.并与化学分析方法进行了比对,两种方法测定值吻合较好. The calibration curve was prepared by the matrix matching method and La was used as the internal standard for calibration to eliminate the influence of matrix effect and instrument drift.The determination method of Al,Mg contents in zinc-aluminum-magnesium alloy by inductively coupled plasma optical emission spectrometry(ICP-OES)was set up.The method was applied to the determination of zinc-aluminum-magnesium alloy sample,the relatively standard deviations were between 0.86% and 2.13%,and the recoveries were between 96.4% and 99.6%.Compared with the chemical analysis method,the results were in good agreement.
作者 常守勤 冯秀梅 陆叶青 陈连芳 王湘静 CHANG Shou-qin;FENG Xiu-mei;LU Ye-qing;CHEN Lian-fang;WANG Xiang-jing(Jiangsu Chengxin Inspection Testing and Certification Co.,Ltd,Jiangyin 214434,Jiangsu China;National Ship Materials Quality Supervision and Testing Center(Jiangsu),Jiangyin 214434,Jiangsu China)
出处 《分析测试技术与仪器》 CAS 2021年第3期209-213,共5页 Analysis and Testing Technology and Instruments
关键词 锌铝镁合金 电感耦合等离子体发射光谱法 zinc-aluminum-magnesium alloy magnesium aluminum ICP-OES
  • 相关文献

参考文献8

二级参考文献71

共引文献92

同被引文献10

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部