Despite the widespread application of point-of-care lateral flow tests,the viscosity dependence of these assay results remains a significant challenge.Here,we employ centrifugal microfluidic flow control through the nitrocellulose membrane of the strip to eliminate the viscosity bias.The key feature is the balancing of the sample flow into the cassette of the lateral flow test with the air flow out of the cassette.A viscosity-indepen3ent flow rate of 3.01±0.18 min(±6%)is demonstrated for samples with viscosities ranging from 1.1 mPas to 24 mPas,a factor greater than 20.In a model human IgG lateral flow assay,signal-intensity shifts caused by varying the sample viscosity from 1.1 mPas to 23 mPas could be reduced by more than 84%.