Flow-induced vibration characteristics of the U-type Coriolis mass flowmeter with liquid hydrogen 被引量:2

原文传递
导出
摘要 Compared with liquid nitrogen(LN_(2))and water,the density of liquid hydrogen(LH_(2))is more than one order of magnitude smaller,which leads to significantly different flow-induced vibration characteristics in the coriolis mass flowmeter(CMF).Based on the Euler beam theory,the complex set of equations of fluid-solid interactions for the U-type pipe Coriolis flowmeter with LH_(2)is solved.The calculation results are firstly validated by comparing the dimensionless frequency,displacement,and twist mode shape with the theoretical and experimental results in the other publications with water and kerosene as the working fluids.Then,the results of dimensionless frequency,phase difference,and time lag for LH_(2)are compared with those for LN_(2)and water,and the effects of the dimensionless flow velocity,sensor position,and the radius of the curved pipe are analyzed in detail for LH_(2).Results show that the time lag of LH_(2)is an order of magnitude smaller than that for LN_(2)or water.The excitation frequency for LH_(2)is much larger than that for LN_(2).Effects of geometric parameters on the time lag are also analyzed for the three fluids and the results contribute to the design optimization of a CMF for LH_(2).
出处 《Journal of Zhejiang University-Science A(Applied Physics & Engineering)》 SCIE EI CAS CSCD 2022年第6期495-504,共10页 浙江大学学报(英文版)A辑(应用物理与工程)
基金 the Key R&D Plan Project of Zhejiang Province(Nos.2021C01099 and 2020C01029),China。