期刊文献+

测氡仪标准仪器校准法的可行性研究 被引量:1

Feasibilit Study on the Calibration Method of Standard Instrument for Emanometer
下载PDF
导出
摘要 针对目前台站测氡仪校准困难的问题,提出标准仪器校准法,并将其与RN-FD循环式氡气固体源进行对比,同时利用台站应用最为广泛的RN-150小型氡气固体源验证该方法的可行性。结果表明,利用RN-150小型氡气固体源及AlphaGUARD P2000F标准仪器分别对FD-125测氡仪3个闪烁室进行校准后得到的校准系数相对误差分别为0.2%、3.4%、2.8%,均达到目前地震监测中氡观测的技术要求,进一步验证了测氡仪标准仪器校准法的可行性。 In view of emanometer calibration difficulty present in earthquake stations,we advance a standard instrument calibration method for emanometer,compare it with RN-FD circulating radon gas solid source,and use RN-150 small radon gas solid source,which is the most widely used gas in current stations,to verify the feasibility of this method.The results show that the three scintillation chambers of FD-125 emanometer are calibrated by using RN-150 small radon gas solid source and AlphaGUARD P2000F emanometer;the relative deviations of the calibration coefficient are 0.2%,3.4%and 2.8%,meeting the technical requirements of radon observation in seismic monitoring,verifing that the standard instrument calibration method is feasible.
作者 任宏微 姚玉霞 陈其峰 REN Hongwei;YAO Yuxia;CHEN Qifeng(Institute of Geophysics,CEA,5 South-Minzudaxue Road,Beijing 100081,China;National Institute of Natural Hazards,MEM,1 Anningzhuang Road,Beijing 100085,China;Pingliang Earthquake Monitoring Center of Gansu Earthquake Agency,231 Mid-Guangcheng Road,Pingliang 744000,China;Liaocheng Earthquake Monitoring Center of Shandong Earthquake Agency,24 South-Zhonghua Road,Liaocheng 252000,China)
出处 《大地测量与地球动力学》 CSCD 北大核心 2023年第1期100-105,共6页 Journal of Geodesy and Geodynamics
基金 国家重点研发计划(2019YFC1509203) 中央级公益性科研院所基本科研业务费专项(ZDJ2021-11) 中国地震局地震科技星火计划(XH21017Y)。
关键词 FD-125测氡仪 AlphaGUARD P2000F测氡仪 闪烁室 标准仪器 RN-150氡气固体源 FD-125 emanometer AlphaGUARD P2000F emanometer scintillation chamber standard instrument RN-150 radon gas solid source
  • 相关文献

参考文献21

二级参考文献165

共引文献106

同被引文献16

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部