摘要
随着新生儿的医疗护理水平不断提高,新生儿监护室的病死率明显降低。由于新生儿免疫功能不全,极易受环境的影响而引发感染,所以预防新生儿医院感染仍然是新生儿病房的首要任务。新生儿暖箱作为大部分新生儿出生后第一时间居住的“家”,已成为新生儿监护室内使用率最高的仪器设备之一,同时也是新生儿监护室基础设备之一。暖箱目前分为密闭式暖箱和长颈鹿多功能培养箱。在半年一次的预防性维护中发现本院有两台暖箱存在运行故障,其中一台长颈鹿多功能培养箱表现为噪音较大,另一台密闭式暖箱表现为湿化故障,湿度无法上升。文章通过对故障现象进行分析,查找故障原因,并采用替换法修复设备,使其恢复到正常工作状态。
With the continuous improvement of medical care of newborns,the case fatality rate in neonatal care unit has decreased significantly.Because the newborn immune function is not complete,it is easy to be affected by the environment and cause infection.Therefore,the prevention of nosocomial infections in neonates remains the top priority of neonatal units.As the“home”where most newborns live for the first time after birth,neonatal warm box has become one of the most frequently used instruments and equipment in neonatal care unit,and also one of the most basic equipment in neonatal care unit.Warm box is currently divided into closed warm box and giraffe multifunctional incubator.During the semi-annual preventive maintenance,we found that two warming tanks in our hospital had operation faults,one giraffe multifunctional incubator showed loud noise,the other closed warming tank showed humidification failure,and the humidity could not rise.This paper mainly describes the fault phenomenon analysis,find the cause of the fault,and use the replacement method to repair the equipment,and make it back to the normal working state.
作者
阮明
夏燕红
RUAN Ming;XIA Yan-hong(Hangzhou Children’s Hospital,Zhejiang Hanghzou 310014)
出处
《中国医疗器械信息》
2024年第5期171-174,共4页
China Medical Device Information
关键词
暖箱检测仪
温度
轴承
风扇
湿化器
水箱
替换法
warm box detector
temperature
bearing
a fan
humidifier
water tank
substitution method