摘要 目的 探讨动脉导管未闭 (PDA)封堵术后溶血的发生原因和处理方法。方法  3例患者 ,PDA最窄处直径分别为 7.5mm、10mm和 11mm ,经静脉途径置入 10 /8、16/14mm直径AmplatzerPDA封堵器和 18mm直径Amplatzer房间隔封堵器。结果  3例患者分别于术后 17h、18h和 5 5min发生溶血。分别采用手术、药物治疗及取出封堵器的方法 ,患者溶血均消失。结论 PDA封堵术后溶血的发生与术后残余分流有关 ,及时有效地处理后疗效满意。 Objective To evaluate the cause and treatments of hemolysis after occlusion of patent ductus arteriosus (PDA). Methods Three patients suffered from PDA with the minimum diameter being 7.5,10 and 11mm.Two Amplatzer PDA occluders (10/8,16/14mm) and an 18mm ASD Amplatzer were implanted by transvenous route. Results In each 17h,18h and 55min after occlusion,the patients developed hemolysis.The hemolysis was resolved after surgery,conservative therapy and retrieval of the device respectively. Conclusion Hemolysis is strongly related to the occurrence of residue after PDA occlusion.With appropriate treatments,the prognosis of hemolysis is satisfatory.
出处 《中国医学影像技术》 CSCD 2003年第2期211-212,共2页 Chinese Journal of Medical Imaging Technology
关键词 动脉导管未闭 封堵术 溶血 疗效 影像学诊断 病例报告 Patent ductus arteriosus Occlusion,interventional Hemolysis
  • 相关文献

参考文献3

  • 1蒋世良 黄连军 等.心血管病介入治疗的严重并发症分析及其防治[J].中国循环杂志,2001,16:22-24.
  • 2陈灏珠.实用内科学[M].北京:人民卫生出版社,1997.858.
  • 3金敬琳 赵世华 蒋世良.Amplatzer封堵器封堵动脉导管未闭术后溶血1例[J].中国医学影像技术,2001,17(2):188-188.