手术前后OrbscanⅡ系统测量误差的分析与校正 被引量:1

The analysis and correction of orbscan Ⅱ topography system's testing disorder
下载PDF
导出
摘要 目的 探讨orbscanⅡ的测量误差及其影响因素 ,为临床使用提供可靠依据。方法 测量lasik 113例 2 2 6眼及lasek14例 2 7眼 ,使用orbscanⅡ系统和A超角膜测厚仪 ,分别于手术前及手术后 1月测量角膜厚度、角膜曲率及角膜后表面屈光力。并对测量值进行分析。结果 lasik组和lasek组 ,OrbscanⅡ术前测量的角膜厚度均较A超值高 ,而术后均较A超值低 ;术后lasik组和lasek组的角膜后表面屈光力较术前均有增加。结论 OrbscanⅡ测量的准确性较高 ,但对于角膜较薄的患眼 。 ObjectiveTo discuss the measurement error and t he influence factor of Orbscan topography to provide more accurate data for clin ical application.MethodsOrbscan topography system and A-scan ultrasonography w ere used to measure the corneal thickness and the curvature of posterior corneal surface.Including 113 cases (226 eyes) undergoing laser in situ keratomileusis (Lasik) and 14 cases (27eyes) undergoing laser epithelial keratomileusis (Lasek) before and after the operations.And the result of measurement was compared.Resultspre-operatively,the value of corneal thickness was low er measured by A-scan ultraconography than by orbscan topography.And post-oper atively the value was lower than that by orbscan topography.The value of posteri or corneal surface curvature after lasik and lasek was more than before.ConclusionThe values obtained by Orbscan topography were accur acy,but there were necessary to improve the accuracy of orbscan topography syste m when we measured the thicker cornea.
出处 《眼外伤职业眼病杂志》 北大核心 2004年第1期4-6,共3页 Journal of Injuries and Occupational Diseases of the Eye with Ophthalmic Surgeries
关键词 角膜地形图 角膜厚度 角膜后表面屈率 Orbscan topography system corneal thickness cur vature of posterior corneal surface
  • 相关文献

参考文献1

  • 1Yaglali V, Kaufman SC. Thompson HW. Corneal thickness measurements with the Orbscan topography system and ultrasonic pachymetry [ J ]. J Cataract Refract Surg, 1997,23:1345 - 1350.

同被引文献6

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部