Aim Portable devices that accurately detect the composition of expired gases and changes in VO_(2)open new possibilities in research methodology and accessibility.Purpose The purpose of this study was to compare the o...Aim Portable devices that accurately detect the composition of expired gases and changes in VO_(2)open new possibilities in research methodology and accessibility.Purpose The purpose of this study was to compare the oxygen consumption(VO_(2))measurements of the VO_(2)Master Pro(VM)to the Cosmed K5(K5)during walking,jogging,and running in field and lab conditions.Methods Twelve proficient runners,with a current 10 k pace of≥11.29 km/h,performed 3 matched intervals at 3 different speeds(4.82,8.05,11.29 km/h)on a treadmill and on an outdoor track.An airflow test was also performed on both devices by pumping air through the devices using a 3 L syringe timed to a metronome at 15,25,and 35 beats/min.Results The VM did not report walking data for most participants.During treadmill running,there were significant differences in VO_(2)(47.86±3.94 vs.29.56±4.15 mL/kg/min),ventilation(V_(e))(71 vs.57 mL/min),and tidal volume(TV)(1.89 vs.1.56 L)between the K5 and VM respectively(P<0.05).Outdoor analysis also showed significant differences between devices in VO_(2),V_(e),and TV(P<0.05).The airflow test showed significant differences between the devices in V_(e)and TV(P<0.05).Conclusion These results suggest that there are significant discrepancies between the K5 and the VM,likely due to differ-ences in TV measurement.展开更多
文摘Aim Portable devices that accurately detect the composition of expired gases and changes in VO_(2)open new possibilities in research methodology and accessibility.Purpose The purpose of this study was to compare the oxygen consumption(VO_(2))measurements of the VO_(2)Master Pro(VM)to the Cosmed K5(K5)during walking,jogging,and running in field and lab conditions.Methods Twelve proficient runners,with a current 10 k pace of≥11.29 km/h,performed 3 matched intervals at 3 different speeds(4.82,8.05,11.29 km/h)on a treadmill and on an outdoor track.An airflow test was also performed on both devices by pumping air through the devices using a 3 L syringe timed to a metronome at 15,25,and 35 beats/min.Results The VM did not report walking data for most participants.During treadmill running,there were significant differences in VO_(2)(47.86±3.94 vs.29.56±4.15 mL/kg/min),ventilation(V_(e))(71 vs.57 mL/min),and tidal volume(TV)(1.89 vs.1.56 L)between the K5 and VM respectively(P<0.05).Outdoor analysis also showed significant differences between devices in VO_(2),V_(e),and TV(P<0.05).The airflow test showed significant differences between the devices in V_(e)and TV(P<0.05).Conclusion These results suggest that there are significant discrepancies between the K5 and the VM,likely due to differ-ences in TV measurement.