Measurement of trace nitrate concentrations in seawater by ion chromatography with valve switching 被引量:2
1
作者 杜娟 法芸 +3 位作者 郑岳 李学兵 杜芳林 杨海燕 《Chinese Journal of Oceanology and Limnology》 SCIE CAS CSCD 2014年第3期732-736,共5页
An ion chromatographic method with a valve switching facility was developed to determine trace nitrate concentrations in seawater using two pumps, two different suppressors, and two columns. A carbohydrate membrane de... An ion chromatographic method with a valve switching facility was developed to determine trace nitrate concentrations in seawater using two pumps, two different suppressors, and two columns. A carbohydrate membrane desalter was used to reduce the high concentrations of sodium salts in samples. In this method, trace nitrate was eluted from the concentrator column to the analytical columns, while the matrix fl owed to waste. Neither chemical pre-treatment nor sample dilution was required. In the optimized separation conditions, the method showed good linearity( R >0.99) in the 0.05 and 50 mg/L concentration range, and satisfactory repeatability(RSD<5%, n =6). The limit of detection for nitrate was 0.02 mg/L. Results showed that the valve switching system was suitable and practical for the determination of trace nitrate in seawater. 展开更多
关键词 硝酸盐浓度 离子色谱法 阀切换 海水 测量 样品稀释 化学预处理 开关设备